Mikster DLM-080 Manual de usuario Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 48
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
The DLM-080 FUTURE Digital Recorder – User’s Manual
26.11.2004
38
Function 67 (Saving Setup Parameters)
Master >>> Slave
Byte
number
Value / variable Range Definition
1 ADDEV
0..31 Slave address in the rs485 network
2 67 (43h)
0..255 Function number
3
.
.
.
130
Setup_Reg
0..255 Block of setup registers transmitted
from
MASTER to SLAVE, length: 128 bytes.
Data format same as SETUP area in the
memory
map.
131 CRC_h
0..255
132 CRC_l 0..255
CRC-16 control total
Function 68 (Clearing of Recording Buffer)
Master >>> Slave
Byte
number
Value / variable Range Definition
1 ADDEV
0..31 Slave address in the rs485 network
2 68 (44h)
0..255 Function number
3 CRC_h
0..255
4 CRC_l 0..255
CRC-16 control total
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 47 48

Comentarios a estos manuales

Sin comentarios